Instrument Introduction
Transmission Electron Microscope (TEM) JEOL JEM-F200
Specifications:
Electron Gun: Cold Field Emiistion Gun(CFEG)
Specimen Holder: Single Tilt Holder; Double Tilt Holder; Three Sample Holder
EDS: JED2300T Super X-Ray System
Supporting Pre-processing System: Zeiss FIB, LeicaTXP, UC7FC7, RES102
Accelerating Voltage: 20-200kV(Standard 80kV, 200kV)
TEM Lattice Resolution:200kV 0.1nm; 80kV 0.14 nm
TEM Specimen Drift: <1nm/min
STEM HAADF Resolution: 200kV 0.136nm; 80kV 0.31nm
STEM BF Resolution:200kV 0.136nm; 80kV 0.31nm
EDS Resolution: <129eV
Function:
TEM, HRTEM, STEM, HAADF, SAED, EDS test for the different kind of sample. At least one dimession of the sample should be lower than 200 nm. "
Dual Beam Electron Microscope (FIB-SEM) Zeiss Cross Beam 350
Specifications:
Electron Gun: Thermal Electron Gun(TFEG)
Ion Source: Ga-Liquid Metal Ion Source
Detector: Chamber SE-detector and In-Lens detector for second electron imaging, aBSD1 and EsB back scattering detector
GIS: Pt and Carbon
FIB resolution at 30kV: 3.0 nm
Second Electron resolution: 1.0 nm@15KV,2.0 nm@1kV
EDS: Analytical silicon drift electric refrigeration detector for the detecting of Be4~Cf98
Cryo system: Quorum PP3005, low temperature to 190℃
Applications:
Second electron imapping, Back scattering electron imapping, X-Ray Analysis, In situ cross-section analysis, TEM sample preparation, 3D construction, Micro and Nano struction processing, 3DAP sample preparation, Cryo FIB, Cryo SEM"
Cryo-Ultramicrotomes Cryo-Ultramicrotomes Leica EM UC7/FC7
Specifications:
Knives: Diatome ultramicrotomy; Diatome cryo diamond knife; Diatome trim diamond knife
Knife making system: EM KMR3 for glass knife making
Section Thickness: 1nm-15μm
Section size: 0.2-14 mm
Speed: 0.05-100mm/s
Working Temperature: -15℃- -185℃
Applications:
High quality sectioning of specimens for light, electron, and atomic force microscopy examination.
Gas Chromatography/Quadrupole-Time of Flight Mass Spectrometry
GC/Q-TOF MS Agilent 8890-7250
Specifications:
Retention time reproducibility:<0.008%
Ion Source:Low Energy EI
Quadrupole Isolation Mass Range:m/z 20-1050
TOF Mass Accuracy :<2ppm for 1 pg OFN at m/z 271.9867
TOF Mass Range:m/z 20-3000
TOF Mass Resolution (FWHM): > 25,000 at m/z 271.9867 independent of acquisition rate
Typical applications:
Target quantification,suspicious screening and non-targeted screening for food and flavoring analysis(Nitrosamine carcinogens, etc),environmental analysis(Polycyclic aromatic hydrocarbons, pesticides, etc),industrial chemicals analysis(chlorinated hydrocarbons, etc),Oil industry analysis(Gasoline, volatile sulfur-containing compounds, etc) and pharmaceutical analysis."
X-ray diffractometer
XRD PANalytical Empyrean powder diffractometer
The anode material is Cu, with K-Alpha1, K-Alpha1, and K-Beta values of 1.54060, 1.54443, and 1.39225 Å. The current and voltage during the test are 40 mA and 45 kV.
Scanning angle range: -6 °~163 °, scanning speed: 0.1 °~50 °/min; Fixed Anti-scatter Slits: 1 °, 1/2 °, 1/4 °, 1/8 °, 1/16 °.
Sample changer 15 positions .
Beta-filter Ni.
Stage flat samples.
Reflection-transmission spinner.
Confocal laser scanning microscope
CLSM Olympus,FV3000
Used for morphology and fluorescence quantitative analysis of 2D/3D (XYZ)/4D (XYZT) for non-destructive analysis of tissues, cells, bacteria, and thin sections。
Laser:405nm、488nm、561nm、640nm
Objective:1.25X、10X、20X、40X、60XO、100XO
Liquid Chromatography Mass Spectrometry LCMS
LC:ThermoFisher Vanquish/ThermoFisher UltiMate 3000 RSLCnano
MS:ThermoFisher Orbitrap Exploris 240
Resolution powers:240000
Mass range:m/z40-6000
Scan speed:30Hz
Typical applications:Biopharmaceutical characterization, proteomics, metabolomics, lipidomics, food and environmental research, clinical research"
Isotope ratio mass spectrometer IRMS
Thermo Scientific™ DELTA V Advantage IRMS
Thermo Scientific™ DELTA V Advantage IRMS
Its sensitivity for CO2 in Dual Inlet mode is better than 1200 molecules per ion. The standard collector configuration is the Universal Triple Collector. H2 collectors with on-line hydrogen capability are optional. The DELTA V Advantage delivers high performance in Dual Inlet and Continuous Flow mode. Excellent linearity is guaranteed with or without helium carrier gas flow.
Connecting the Elemental Analyzer FLASH IRMS to an Isotope Ratio Mass Spectrometer allows for two-element analysis of C and N or H and O with a dynamic range greater than 1:1000.
The Thermo Scientific GC IsoLink TRACE 1310with combustion and high temperature pyrolysis is used for compound specific isotope analysis of complex mixtures separated on a capillary GC column. Determination of D/H, 13C/12C, 15N/14N and 18O/16O isotope ratios is possible for all mixtures accessible by the GC technique.
IRMS is used with GasBench II foratmospheric trace-gas research in the ppm and ppb range.
Flow cytometer FCM BECKMAN COULTER CytoFLEX
Laser:375nm、405nm、488nm、638nm、561nm、808nm
Minimum sample volume:≤10ul
Dynamic linear range of the signal:≥7-decade dynamic range
Speed of analysis:≥3000 cell/sec
Fluorescence sensitivity:FITC≤30MESF、PE≤10MESF
Typical applications:Immunophenotyping、Cell Sorting、Cell Cycle Analysis、Apoptosis、Cell Proliferation Assays、Intracellular Calcium Flux"
High resolution stable isotope mass spectrometry
Panorama NU AMETEK Panorama
The Panorama, takes a completely new ion optical approach for these type of instruments, and is based on the ESA – Quadrupole – Magnet design of Matsuda. Optimised for ultra-high resolution studies, whilst maintaining high sensitivity, this design uses a magnet of 800mm radius, and has been shown to resolve all the methane peaks at mass 18amu, to permit the first truly sub permille measurement of, not only 13CH3D, but also 12CH2D2. The mass dispersion / magnification figure of merit for this instrument is 1400mm.
X-ray Photoelectron Spectroscopy
XPS Brand:PHI
Model:PHI Versaprobe 4
Function:Dual beam charge neutralization;Argon ion gun, cluster ion gun ;Inert atmosphere transfer tubes.etc.XPS is widely used to analyze qualitative and quantitative analysis of inorganic compounds, alloys, semiconductors, polymers, elements, catalysts, glass, ceramics, dyes, paper, inks, wood, cosmetics, biomaterials, compound valence analysis, in-depth analysis of materials."
X-ray Microscopy/Computed Tomography
XRM/ X-ray CT ZEISS Xradia 610 Versa
higest Resolution: 0.5 μm
X-ray Source: Voltage Range: 30 – 160 kV; Maximum Output: 25 W
Objectives: 0.4x, 4x, 20x
The 3D X-ray microscope enables non-destructive imaging of internal structures with high resolution. It offers detailed external and internal 3D imaging, allowing for precise analysis of materials, defects, and microstructures in various fields such as materials science, biology, and geology etc."
Nano particle size and zeta potential analyzer
DLS Brand:Anton Paar
Model:Litesizer 500
Function:The basic principle of dynamic light scattering measurement of nanoparticles is based on the Brownian motion of nanoparticles. Smaller particles move at a higher speed than larger particles. The hydrodynamic diameter of the particles can be determined by measuring the speed of the particles."
Gas Adsorption Analyzer
BET Brand:Anton Paar
Model:Autosorb-iQ
Function:For key substances involved in the fields of drying, catalysis, liquid purification, gas storage, respiratory protection, pollution control, etc., physical adsorption is the core to achieve their functionality. To achieve these functions, these materials need to have corresponding characteristics in terms of structural characteristics, including specific surface area, porosity, pore size, pore volume, adsorption heat, adsorption kinetics, etc. Among them, specific surface area, pore size, pore volume, etc. are the most commonly used characterization parameters"
Continuous flow analyzer
CFA Brand:Seal
Model:AA 500
Function:Detection of phosphate, ammonium salt, nitrate, nitrite and silicate in water, wastewater, plants, soil extracts, seawater, tobacco, feed, and fertilizers."
Nano Secondary Ion Mass Spectrometer NanoSIMS
Brand:Cameca Model:NanoSIMS 50L
The NanoSIMS is a unique ion microprobe that allows the analysis of elemental and isotopic composition of a solid sample at a sub-micron spatial resolution. It delivers simultaneously key performance metrics include high analysis spatial resolution, high transmission, and high sensitivity. It is mainly used in astrochemistry, geochemistry, microbiology, cell biology, and materials science. It provides information about the lateral distribution of elements and its isotopes as well as the quantitative information about the isotopic composition of a sample."
High pressure freezing & freezing substitution system
HPF & FS Brand: Leica
Model:EM ICE, EM AFS2
High pressure freezing is key for the study of intricate changes in fine structure or cellular dynamics. Currently cryo-fixation is the only way to fix cellular constituents without introducing significant structural alterations. It can cryo-immobilize aqueous samples under high pressure with a unique freezing principle and uncover secrets of the cellular process. The EM ICE solution uses a unique, alcohol-free freezing principle to allow a superior cryo-fixation of the specimen enabling better quality results to be obtained."
Multi-collectoer Inductively coupled plasma mass spectrometry
MC-ICP-MS Brand: Thermo Fisher Scientific
Modle: Neptune XT
Function: MC-ICP-MS combine the Ar ICP-source of conventional ICP-MS instruments with the magnetic sector analyser and multiple-Faraday cup array of TIMS. It can be used to analyze the isotope ratios of most elements in the periodic table, such as lithium, strontium, copper and Zinc."
laser ablation system LA Brand: Elemental Scientific Lasers
Modle: NWR193HE
Function: The laser ablation system is a widely used analytical technique for for in-situ micro analysis of solid samples, combined with ICP-MS to determine element content, and combined with MC-ICP-MS to determine the isotopic composition. High sensitivity, good dynamic range of more than ten orders of magnitude and minimal sample preparation made this technique popular in environment, biology and particularly geosciences."
Ion Chromatography Inductively Coupled Plasma Mass Spectrometry IC-ICP-MS
IC: Themo Fisher ICS 6000
Ion Chromatography (IC) is a powerful analytical technique used for separating and quantifying ions in a sample.
ICP MS: Themo Fisher iCAP RQ
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) is an extremely sensitive analytical technique used for detecting and quantifying trace elements at very low concentrations, typically in the parts-per-billion (ppb) or parts-per-trillion (ppt) range.
IC-ICP-MS is a powerful hyphenated technique that combines Ion Chromatography (IC) with Inductively Coupled Plasma Mass Spectrometry (ICP-MS).